The recommended maximum tolerance of kVp variance for mammographic x-ray generators is

A) 1%
B) 2%
C) 5%
D) 10%
استخدم زر المسافة أو
up arrow
down arrow
لقلب البطاقة.
سؤال
The B + F value determined from sensitometric testing of mammographic film processors must not vary by more than ____ from the initial control value.

A) ±\pm 0.01%
B) ±\pm 0.03%
C) ±\pm 0.05%
D) ±\pm 100.10%
سؤال
The contrast indicator, or density difference, value from sensitometric testing of mammographic film processors must not vary by more than ____ from the initial control value.

A) ±\pm 0.01%
B) ±\pm 0.03%
C) ±\pm 0.05%
D) ±\pm 0.15%
سؤال
The average glandular dose delivered during a single craniocaudal view of an FDA-accepted phantom simulating a standard breast shall not exceed ____ mGy per exposure.

A) 1
B) 2
C) 3
D) 5
سؤال
The MQSA does not require facilities to have

A) annual physics testing.
B) minimum image quality standards.
C) ACR accreditation.
D) FDA certification.
سؤال
The background density obtained during phantom image testing should not vary by more than _____ from the established operating level.

A) ±0.2
B) ±0.3
C) ±0.4
D) ±0.5
سؤال
Mammographers must perform phantom image evaluation at least

A) daily.
B) weekly.
C) monthly.
D) semi-annually.
سؤال
The mAs recorded during phantom image testing of mammographic equipment should not vary by more than ____ from the initial control value.

A) ±\pm 2%
B) ±\pm 5%
C) ±\pm 10%
D) ±\pm 15%
سؤال
Developer temperature in mammographic film processors should not vary by more than

A) ±\pm 0.5 °\degree F.
B) ±\pm 1.0 °\degree F.
C) ±\pm 2.0 °\degree F.
D) ±\pm 5.0 °\degree F.
سؤال
The maximum allowable dose per exposure for mammographic images is ____ rad.

A) 0.1
B) 0.2
C) 0.3
D) 0.5
فتح الحزمة
قم بالتسجيل لفتح البطاقات في هذه المجموعة!
Unlock Deck
Unlock Deck
1/10
auto play flashcards
العب
simple tutorial
ملء الشاشة (f)
exit full mode
Deck 9: Mammographic Quality Standards
1
The recommended maximum tolerance of kVp variance for mammographic x-ray generators is

A) 1%
B) 2%
C) 5%
D) 10%
5%
2
The B + F value determined from sensitometric testing of mammographic film processors must not vary by more than ____ from the initial control value.

A) ±\pm 0.01%
B) ±\pm 0.03%
C) ±\pm 0.05%
D) ±\pm 100.10%
±\pm 0.03%
3
The contrast indicator, or density difference, value from sensitometric testing of mammographic film processors must not vary by more than ____ from the initial control value.

A) ±\pm 0.01%
B) ±\pm 0.03%
C) ±\pm 0.05%
D) ±\pm 0.15%
±\pm 0.15%
4
The average glandular dose delivered during a single craniocaudal view of an FDA-accepted phantom simulating a standard breast shall not exceed ____ mGy per exposure.

A) 1
B) 2
C) 3
D) 5
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
5
The MQSA does not require facilities to have

A) annual physics testing.
B) minimum image quality standards.
C) ACR accreditation.
D) FDA certification.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
6
The background density obtained during phantom image testing should not vary by more than _____ from the established operating level.

A) ±0.2
B) ±0.3
C) ±0.4
D) ±0.5
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
7
Mammographers must perform phantom image evaluation at least

A) daily.
B) weekly.
C) monthly.
D) semi-annually.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
8
The mAs recorded during phantom image testing of mammographic equipment should not vary by more than ____ from the initial control value.

A) ±\pm 2%
B) ±\pm 5%
C) ±\pm 10%
D) ±\pm 15%
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
9
Developer temperature in mammographic film processors should not vary by more than

A) ±\pm 0.5 °\degree F.
B) ±\pm 1.0 °\degree F.
C) ±\pm 2.0 °\degree F.
D) ±\pm 5.0 °\degree F.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
10
The maximum allowable dose per exposure for mammographic images is ____ rad.

A) 0.1
B) 0.2
C) 0.3
D) 0.5
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
locked card icon
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.