Deck 7: Quality Control of Fluoroscopic Equipment

ملء الشاشة (f)
exit full mode
سؤال
At any combination of tube potential and current, fluoroscopic equipment with an automatic brightness control should not produce an entrance skin exposure exceeding ____ R/min.

A) 2.5
B) 5
C) 10
D) 20
استخدم زر المسافة أو
up arrow
down arrow
لقلب البطاقة.
سؤال
A photocathode is a device that emits

A) light photons.
B) x-ray photons.
C) electrons.
D) All of the answer options are correct.
سؤال
Which type of distortion that can occur with tube-type fluoroscopic image intensifiers is most often caused by the presence of strong magnetic fields?

A) Veiling glare
B) S distortion
C) Pincushion distortion
D) Vignetting
سؤال
The input phosphor of an image intensifier tube converts a(n) _____ image into a(n) _____ image.

A) x-ray; light
B) light; electron
C) electron; light
D) x-ray; electron
سؤال
The television camera most commonly used to capture the fluoroscopic image from a tube type image intensifier is the:

A) image orthicon.
B) plumbicon.
C) vidicon.
D) charge-coupled device.
سؤال
A reduction in brightness at the periphery of the image captured with an image intensifier tube is referred to as

A) pincushion distortion.
B) vignetting.
C) barrel distortion.
D) quantum mottle.
سؤال
Performance testing of fluoroscopic systems must include all the following except

A) x-ray tube housing.
B) automatic film processor.
C) total filtration.
D) automatic brightness control.
سؤال
Quality control tests of fluoroscopic systems should be performed

A) daily.
B) weekly.
C) semiannually.
D) annually.
سؤال
Which type of test tool may be used to demonstrate pincushion distortion?

A) Dosimeter
B) Wire mesh tool
C) Ion chamber
D) Homogenous phantom
سؤال
In standard broadcast television, one image frame consists of

A) one television field.
B) two television fields.
C) 262.5 scan lines.
D) 17 ms.
فتح الحزمة
قم بالتسجيل لفتح البطاقات في هذه المجموعة!
Unlock Deck
Unlock Deck
1/10
auto play flashcards
العب
simple tutorial
ملء الشاشة (f)
exit full mode
Deck 7: Quality Control of Fluoroscopic Equipment
1
At any combination of tube potential and current, fluoroscopic equipment with an automatic brightness control should not produce an entrance skin exposure exceeding ____ R/min.

A) 2.5
B) 5
C) 10
D) 20
C
The intensity of the x-ray beam at tabletop should not exceed 10 R/min (2.6 mC kg-1min-1) for units equipped with ABC.
2
A photocathode is a device that emits

A) light photons.
B) x-ray photons.
C) electrons.
D) All of the answer options are correct.
C
The light photons release electrons from the photocathode through the process of photoemission.
3
Which type of distortion that can occur with tube-type fluoroscopic image intensifiers is most often caused by the presence of strong magnetic fields?

A) Veiling glare
B) S distortion
C) Pincushion distortion
D) Vignetting
B
An S distortion artifact is a warping of the image along an S-shaped axis and is the result of strong magnetic fields changing the trajectory of the electrons moving across the image intensifier tube.
4
The input phosphor of an image intensifier tube converts a(n) _____ image into a(n) _____ image.

A) x-ray; light
B) light; electron
C) electron; light
D) x-ray; electron
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
5
The television camera most commonly used to capture the fluoroscopic image from a tube type image intensifier is the:

A) image orthicon.
B) plumbicon.
C) vidicon.
D) charge-coupled device.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
6
A reduction in brightness at the periphery of the image captured with an image intensifier tube is referred to as

A) pincushion distortion.
B) vignetting.
C) barrel distortion.
D) quantum mottle.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
7
Performance testing of fluoroscopic systems must include all the following except

A) x-ray tube housing.
B) automatic film processor.
C) total filtration.
D) automatic brightness control.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
8
Quality control tests of fluoroscopic systems should be performed

A) daily.
B) weekly.
C) semiannually.
D) annually.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
9
Which type of test tool may be used to demonstrate pincushion distortion?

A) Dosimeter
B) Wire mesh tool
C) Ion chamber
D) Homogenous phantom
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
10
In standard broadcast television, one image frame consists of

A) one television field.
B) two television fields.
C) 262.5 scan lines.
D) 17 ms.