Deck 10: Quality Control in Computed Tomography

ملء الشاشة (f)
exit full mode
سؤال
Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.

A) ±\pm 2%
B) ±\pm 5%
C) ±\pm 10%
D) ±\pm 20%
استخدم زر المسافة أو
up arrow
down arrow
لقلب البطاقة.
سؤال
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.

A) 2
B) 3
C) 5
D) 10
سؤال
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

A) ±\pm 0.5
B) ±\pm 1.0
C) ±\pm 2.0
D) ±\pm 5.0
سؤال
The laser light field and radiation field should coincide to within ____ mm.

A) 0.5
B) 1.0
C) 2.0
D) 5.0
سؤال
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

A) daily.
B) weekly.
C) monthly.
D) semi-annually.
سؤال
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

A) Low-contrast resolution
B) Uniformity
C) High-contrast resolution
D) Image noise
سؤال
The CT number of water should not deviate by more than ____ CT numbers from 0.

A) 2
B) 3
C) 5
D) 10
سؤال
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

A) 0.5
B) 1.0
C) 2.0
D) 3.0
سؤال
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

A) quarterly.
B) monthly.
C) weekly.
D) daily.
سؤال
The CTDIvol pass/fail criterion for an adult abdomen is ____ mGy.

A) 20
B) 40
C) 30
D) 80
فتح الحزمة
قم بالتسجيل لفتح البطاقات في هذه المجموعة!
Unlock Deck
Unlock Deck
1/10
auto play flashcards
العب
simple tutorial
ملء الشاشة (f)
exit full mode
Deck 10: Quality Control in Computed Tomography
1
Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.

A) ±\pm 2%
B) ±\pm 5%
C) ±\pm 10%
D) ±\pm 20%
±\pm 10%
2
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.

A) 2
B) 3
C) 5
D) 10
C
Edge-to-center mean CT number difference must be less than 5 HU for all four edge positions.
3
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

A) ±\pm 0.5
B) ±\pm 1.0
C) ±\pm 2.0
D) ±\pm 5.0
±\pm 1.0
4
The laser light field and radiation field should coincide to within ____ mm.

A) 0.5
B) 1.0
C) 2.0
D) 5.0
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
5
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

A) daily.
B) weekly.
C) monthly.
D) semi-annually.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
6
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

A) Low-contrast resolution
B) Uniformity
C) High-contrast resolution
D) Image noise
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
7
The CT number of water should not deviate by more than ____ CT numbers from 0.

A) 2
B) 3
C) 5
D) 10
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
8
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

A) 0.5
B) 1.0
C) 2.0
D) 3.0
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
9
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

A) quarterly.
B) monthly.
C) weekly.
D) daily.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
10
The CTDIvol pass/fail criterion for an adult abdomen is ____ mGy.

A) 20
B) 40
C) 30
D) 80
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck