Deck 11: Mammographic Quality Standards
1
Developer temperature in mammographic film processors should not vary by more than
A)±0.5° F.
B)±1.0° F.
C)±2.0° F.
D)±5.0° F.
A)±0.5° F.
B)±1.0° F.
C)±2.0° F.
D)±5.0° F.
A
2
The MQSA does not require facilities to have
A)annual physics testing.
B)minimum image quality standards.
C)ACR accreditation.
D)FDA certification.
A)annual physics testing.
B)minimum image quality standards.
C)ACR accreditation.
D)FDA certification.
C
3
Mammographers must perform phantom image evaluation at least
A)daily.
B)weekly.
C)monthly.
D)semi-annually.
A)daily.
B)weekly.
C)monthly.
D)semi-annually.
B
4
The average glandular dose delivered during a single craniocaudal view of an FDA-accepted phantom simulating a standard breast shall not exceed ____ mGy per exposure.
A)1
B)2
C)3
D)5
A)1
B)2
C)3
D)5
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
5
The maximum allowable dose per exposure for mammographic images is ____ rad.
A)0.1
B)0.2
C)0.3
D)0.5
A)0.1
B)0.2
C)0.3
D)0.5
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
6
The mAs recorded during phantom image testing of mammographic equipment should not vary by more than ____ from the initial control value.
A)±2%
B)±5%
C)±10%
D)±15%
A)±2%
B)±5%
C)±10%
D)±15%
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
7
The recommended maximum tolerance of kVp variance for mammographic x-ray generators is
A)1%.
B)2%.
C)5%.
D)10%.
A)1%.
B)2%.
C)5%.
D)10%.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
8
The contrast indicator, or density difference, value from sensitometric testing of mammographic film processors must not vary by more than ____ from the initial control value.
A)±0.01%
B)±0.03%
C)±0.05%
D)±0.15%
A)±0.01%
B)±0.03%
C)±0.05%
D)±0.15%
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
9
The B + F value determined from sensitometric testing of mammographic film processors must not vary by more than ____ from the initial control value.
A)±0.01%
B)±0.03%
C)±0.05%
D)±0.10%
A)±0.01%
B)±0.03%
C)±0.05%
D)±0.10%
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
10
The background density obtained during phantom image testing should not vary by more than _____ from the established operating level.
A)±0.2
B)±0.3
C)±0.4
D)±0.5
A)±0.2
B)±0.3
C)±0.4
D)±0.5
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck