Deck 13: Quality Control for Magnetic Resonance Imaging Equipment
1
What is the resonant frequency in a unit with a 1.0-T magnet and a gyromagnetic ratio of 42.57 MHz/T?
A)12.77 MHz
B)21.29 MHz
C)42.57 MHz
D)63.86 MHz
A)12.77 MHz
B)21.29 MHz
C)42.57 MHz
D)63.86 MHz
C
2
Nonuniformity in the magnetic field can be caused by
A)a homogenous main magnetic field.
B)uniformity in the RF coil.
C)the presence of ferromagnetic material within the bore of the magnet.
D)all of the above.
A)a homogenous main magnetic field.
B)uniformity in the RF coil.
C)the presence of ferromagnetic material within the bore of the magnet.
D)all of the above.
C
3
Luminance uniformity of soft-copy displays used in MRI systems should be within ____ of the maximum brightness measured in the center of the screen.
A)10%
B)20%
C)30%
D)50%
A)10%
B)20%
C)30%
D)50%
C
4
The amount of geometric distortion in the image is the
A)signal-to-noise ratio.
B)image uniformity.
C)spatial resolution.
D)spatial linearity.
A)signal-to-noise ratio.
B)image uniformity.
C)spatial resolution.
D)spatial linearity.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
5
What is the maximum luminance for soft-copy display devices as suggested by the ACR?
A)Less than 1.2 nit
B)Greater than 50 nit
C)Greater than 90 nit
D)Less than 150 nit
A)Less than 1.2 nit
B)Greater than 50 nit
C)Greater than 90 nit
D)Less than 150 nit
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
6
The ability of the MR system to differentiate between adjacent tissues having minimal differences in signal intensities is known as
A)signal-to-noise ratio.
B)high-contrast resolution.
C)low-contrast resolution.
D)spatial linearity.
A)signal-to-noise ratio.
B)high-contrast resolution.
C)low-contrast resolution.
D)spatial linearity.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
7
The ability of an MR scanner to resolve small objects is known as
A)signal-to-noise ratio.
B)high-contrast resolution.
C)spatial resolution.
D)spatial linearity.
A)signal-to-noise ratio.
B)high-contrast resolution.
C)spatial resolution.
D)spatial linearity.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
8
A change in the strength of the main magnetic field can be caused by
A)cryogen boiloff.
B)external ferromagnetic materials.
C)shim coil failure.
D)all of the above.
A)cryogen boiloff.
B)external ferromagnetic materials.
C)shim coil failure.
D)all of the above.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
9
An acquisition with a 200-mm field of view and 256 x and y encodes yields a pixel size of ____ mm².
A)0.78
B)1.56
C)2.08
D)none of the above
A)0.78
B)1.56
C)2.08
D)none of the above
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
10
The ACR slice position deviation in MR scanners should not exceed ____ mm.
A)1
B)2
C)3
D)5
A)1
B)2
C)3
D)5