Deck 25: The Imaging Process

ملء الشاشة (f)
exit full mode
سؤال
If the image quality acceptance limits are narrow,

A)image repeat rates may be high.
B)the department would be a good candidate for students.
C)the radiography staff would not expect to need regular inservices.
D)patient dosages would be unaffected.
استخدم زر المسافة أو
up arrow
down arrow
لقلب البطاقة.
سؤال
Which phase of the imaging process includes control of factors such as density, contrast, recorded detail, and distortion?

A)image display
B)image processing
C)image analysis
D)image acquisition
سؤال
Visibility of detail quality factors include all of the following EXCEPT

A)radiographic image density.
B)x-ray tube focal spot size.
C)image contrast.
D)radiographic grids.
سؤال
The acceptance limit curve is often skewed to the ____ side.

A)low-contrast
B)high-contrast
C)light
D)none of the above
سؤال
Factors that affect the shape of the acceptance limit curve of medical images include all of the following EXCEPT

A)equipment condition.
B)department quality standards.
C)department location.
D)positioning abilities of the radiographer.
سؤال
Geometric image quality factors include:
1)focal spot selection
2)SID
3)grid ratio
4)filtration thickness
5)OID
6)tube angulation
7)part thickness
8)receptor resolution

A)3, 4, 7, and 8 only
B)1, 2, 4, and 7 only
C)1, 2, 5, 6, 7, and 8 only
D)1, 2, 3, 4, and 7 only
سؤال
What process is used to create a hard copy of digital images?

A)laser
B)DICOM formatting
C)wet mounting
D)You can't create hard copies of digital images.
سؤال
After a radiograph of the hip in the translateral projection is completed, the final image is judged by one of the radiographers as marginally acceptable, as it might show the acetabular rim and femoral head.The logical correct action to take would be to

A)present the image to a visiting radiologist who is there for the day.
B)present the image to a radiologist known to be "popular" with technologists.
C)repeat the image to obtain better image quality.
D)send the image for interpretation but make a note of the image in the quality control file of images.
سؤال
A KUB is completed on an obese patient using AEC and 70kVp.The x-ray generator terminated early after hitting the preset backup mAs value.The final image reveals a low EI value.Effective image analysis could reveal that

A)the combination of AEC detectors was incorrect.
B)an increase of 2 kVp is necessary.
C)the backup mAs needs to be recalibrated.
D)an increase in grid ratio is all that is needed.
سؤال
Image quality is primarily the domain of the

A)x-ray service engineer.
B)administrator.
C)radiographer.
D)physicist.
سؤال
Which phase of the imaging process incorporates the concepts of spatial resolution, density and brightness, and contrast and dynamic range?

A)image analysis
B)image acquisition
C)image display
D)image processing
سؤال
Which of these activities would be inconsistent with the "scope of practice" of the radiographer?

A)suggesting methods to improve fluoroscopic image quality to a radiologist
B)helping an ER physician identify a pneumonia on a chest radiograph
C)obtaining additional patient information to add to an existing patient history that was taken by a nurse
D)repeating a suboptimal radiograph after determining the S# was too high
سؤال
After completing a chest radiograph on a female patient with large pendulous breasts, the final image demonstrates acceptable density and contrast but a loss of recorded detail in the lung bases.The professional radiographer would

A)repeat the chest radiograph in the anterio-posterior (AP) projection.
B)increase the SID to improve resolution.
C)increase the mAs and tell the patient to hold their breath deeper.
D)politely ask the patient to separate her breasts as she approaches the chest board and repeat the exam in the PA projection.
سؤال
Producing an acceptable radiograph requires consideration of

A)technical factor selection.
B)pathology.
C)image processing.
D)all of the above
سؤال
When many radiographic factors are acting upon one other, it is known as

A)a pyramid problem.
B)a professional problem.
C)procedural problem.
D)an ethical problem.
سؤال
After a radiograph of the lumbar spine (AP) is completed, the final image reveals a S# of 758.Your correct image analysis leads you to

A)increase the mAs by 10% and repeat the image.
B)increase the kVp by 4 kVp and repeat the image.
C)decrease the SID to 36" and repeat the image.
D)none of the above
سؤال
All of the following are related except one.Select the choice not related to the others.

A)image analysis
B)image acquisition
C)image transmission
D)image processing
سؤال
Upon analysis of a chest radiograph taken on a female patient, the radiographer notices that there is an uneven radiographic density over the lung bases.One lung base is significantly underexposed compared to the other.In an effort to determine the possible cause(s), a logical strategy to take would be to

A)inspect the AEC detectors and retake the image.
B)retake the image on expiration.
C)ask the patient if she has had a previous breast surgery.
D)make a note for the radiologist that you cannot explain the opacification, as the patient is asymptomatic.
سؤال
Which phase of the imaging process is unique to each imaging modality?

A)image analysis
B)image acquisition
C)image archiving
D)image display
سؤال
Generally speaking, the primary cause of increased patient radiation dose is

A)extended fluoroscopy.
B)higher grid ratios than necessary.
C)inadequate beam restriction.
D)repeat images.
سؤال
Match the following choices with the correct statement as they relate to the imaging process.
production of final image

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
سؤال
Match the following choices with the correct statement as they relate to the imaging process.
assessing image quality in terms of density, contrast, detail, distortion

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
سؤال
Match the following choices with the correct statement as they relate to the imaging process.
high image quality

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
سؤال
Match the following choices with the correct statement as they relate to the imaging process.
agreed-upon image quality standards

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
سؤال
Match the following choices with the correct statement as they relate to the imaging process.
first phase in the diagnostic process

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
سؤال
Match the following choices with the correct statement as they relate to the imaging process.
combination of inter-related factors that compound the problem

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
فتح الحزمة
قم بالتسجيل لفتح البطاقات في هذه المجموعة!
Unlock Deck
Unlock Deck
1/26
auto play flashcards
العب
simple tutorial
ملء الشاشة (f)
exit full mode
Deck 25: The Imaging Process
1
If the image quality acceptance limits are narrow,

A)image repeat rates may be high.
B)the department would be a good candidate for students.
C)the radiography staff would not expect to need regular inservices.
D)patient dosages would be unaffected.
image repeat rates may be high.
2
Which phase of the imaging process includes control of factors such as density, contrast, recorded detail, and distortion?

A)image display
B)image processing
C)image analysis
D)image acquisition
image analysis
3
Visibility of detail quality factors include all of the following EXCEPT

A)radiographic image density.
B)x-ray tube focal spot size.
C)image contrast.
D)radiographic grids.
x-ray tube focal spot size.
4
The acceptance limit curve is often skewed to the ____ side.

A)low-contrast
B)high-contrast
C)light
D)none of the above
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
5
Factors that affect the shape of the acceptance limit curve of medical images include all of the following EXCEPT

A)equipment condition.
B)department quality standards.
C)department location.
D)positioning abilities of the radiographer.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
6
Geometric image quality factors include:
1)focal spot selection
2)SID
3)grid ratio
4)filtration thickness
5)OID
6)tube angulation
7)part thickness
8)receptor resolution

A)3, 4, 7, and 8 only
B)1, 2, 4, and 7 only
C)1, 2, 5, 6, 7, and 8 only
D)1, 2, 3, 4, and 7 only
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
7
What process is used to create a hard copy of digital images?

A)laser
B)DICOM formatting
C)wet mounting
D)You can't create hard copies of digital images.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
8
After a radiograph of the hip in the translateral projection is completed, the final image is judged by one of the radiographers as marginally acceptable, as it might show the acetabular rim and femoral head.The logical correct action to take would be to

A)present the image to a visiting radiologist who is there for the day.
B)present the image to a radiologist known to be "popular" with technologists.
C)repeat the image to obtain better image quality.
D)send the image for interpretation but make a note of the image in the quality control file of images.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
9
A KUB is completed on an obese patient using AEC and 70kVp.The x-ray generator terminated early after hitting the preset backup mAs value.The final image reveals a low EI value.Effective image analysis could reveal that

A)the combination of AEC detectors was incorrect.
B)an increase of 2 kVp is necessary.
C)the backup mAs needs to be recalibrated.
D)an increase in grid ratio is all that is needed.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
10
Image quality is primarily the domain of the

A)x-ray service engineer.
B)administrator.
C)radiographer.
D)physicist.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
11
Which phase of the imaging process incorporates the concepts of spatial resolution, density and brightness, and contrast and dynamic range?

A)image analysis
B)image acquisition
C)image display
D)image processing
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
12
Which of these activities would be inconsistent with the "scope of practice" of the radiographer?

A)suggesting methods to improve fluoroscopic image quality to a radiologist
B)helping an ER physician identify a pneumonia on a chest radiograph
C)obtaining additional patient information to add to an existing patient history that was taken by a nurse
D)repeating a suboptimal radiograph after determining the S# was too high
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
13
After completing a chest radiograph on a female patient with large pendulous breasts, the final image demonstrates acceptable density and contrast but a loss of recorded detail in the lung bases.The professional radiographer would

A)repeat the chest radiograph in the anterio-posterior (AP) projection.
B)increase the SID to improve resolution.
C)increase the mAs and tell the patient to hold their breath deeper.
D)politely ask the patient to separate her breasts as she approaches the chest board and repeat the exam in the PA projection.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
14
Producing an acceptable radiograph requires consideration of

A)technical factor selection.
B)pathology.
C)image processing.
D)all of the above
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
15
When many radiographic factors are acting upon one other, it is known as

A)a pyramid problem.
B)a professional problem.
C)procedural problem.
D)an ethical problem.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
16
After a radiograph of the lumbar spine (AP) is completed, the final image reveals a S# of 758.Your correct image analysis leads you to

A)increase the mAs by 10% and repeat the image.
B)increase the kVp by 4 kVp and repeat the image.
C)decrease the SID to 36" and repeat the image.
D)none of the above
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
17
All of the following are related except one.Select the choice not related to the others.

A)image analysis
B)image acquisition
C)image transmission
D)image processing
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
18
Upon analysis of a chest radiograph taken on a female patient, the radiographer notices that there is an uneven radiographic density over the lung bases.One lung base is significantly underexposed compared to the other.In an effort to determine the possible cause(s), a logical strategy to take would be to

A)inspect the AEC detectors and retake the image.
B)retake the image on expiration.
C)ask the patient if she has had a previous breast surgery.
D)make a note for the radiologist that you cannot explain the opacification, as the patient is asymptomatic.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
19
Which phase of the imaging process is unique to each imaging modality?

A)image analysis
B)image acquisition
C)image archiving
D)image display
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
20
Generally speaking, the primary cause of increased patient radiation dose is

A)extended fluoroscopy.
B)higher grid ratios than necessary.
C)inadequate beam restriction.
D)repeat images.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
21
Match the following choices with the correct statement as they relate to the imaging process.
production of final image

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
22
Match the following choices with the correct statement as they relate to the imaging process.
assessing image quality in terms of density, contrast, detail, distortion

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
23
Match the following choices with the correct statement as they relate to the imaging process.
high image quality

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
24
Match the following choices with the correct statement as they relate to the imaging process.
agreed-upon image quality standards

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
25
Match the following choices with the correct statement as they relate to the imaging process.
first phase in the diagnostic process

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
26
Match the following choices with the correct statement as they relate to the imaging process.
combination of inter-related factors that compound the problem

A)narrowing the search
B)pyramid problem
C)image processing
D)wide acceptance limits
E)image acquisition
F)acceptable limits
G)narrow acceptance limits
H)image analysis
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.
فتح الحزمة
k this deck
locked card icon
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 26 في هذه المجموعة.