Deck 16: Automation

ملء الشاشة (f)
exit full mode
سؤال
The type of automated analyzer in which the operator is able to use in-house reagents or reagents purchased from suppliers different from the analyzer's manufacturer is referred to as a(n):

A) discrete analyzer.
B) random-access analyzer.
C) open analyzer system.
D) closed analyzer system.
استخدم زر المسافة أو
up arrow
down arrow
لقلب البطاقة.
سؤال
Your laboratory has purchased a discrete processing analyzer that is dedicated to performing only thyroid hormone analysis.Which type of pipetting system will likely be in place in this instrument?

A) Positive liquid displacement fixed-volume
B) Air displacement selectable-volume
C) Positive liquid displacement selectable-volume
D) Air displacement fixed-volume
سؤال
The most commonly used method of reducing carryover in most automated analytical systems is to use:

A) extensive system flushing between specimens and sample probe washing.
B) plastic reagent containers with closed lids.
C) disposable cuvets.
D) a specific measurement approach that eliminates any possibility of carryover.
سؤال
If programmed appropriately,a multiple-channel analyzer could assess samples as a single-channel analyzer.
سؤال
What component causes carryover of the analyte in a discrete analytical system that uses disposable reaction vessels and measuring cuvets?

A) The point at which sequential samples pass through
B) The pipetting system
C) The peristaltic pump
D) The reagent delivery system
سؤال
A tray of clear sample cups sits on a laboratory bench top for an extended period of time.Which one of the following analytes might be affected?

A) Protein
B) Cholesterol
C) Glucose
D) Bilirubin
سؤال
Using whole blood as the specimen of choice in an automated analytical system:

A) essentially eliminates specimen preparation time.
B) allows the operator to use a secondary tube for analysis.
C) keeps the specimen from undergoing degradation.
D) allows for the avoidance of carryover.
سؤال
An ion-selective electrode system for electrolyte analysis that uses a peristaltic pump to move liquid reagents and samples through a system of tubing is called a _____ analyzer.

A) random access
B) batch
C) discrete
D) continuous flow
سؤال
Specimen bar coding has allowed for the ability of a system operator to place specimens in random sequence,thereby eliminating the need for specimen load lists.
سؤال
The point of interaction between the instrument operator and an automated analyzer that is dedicated to a defined task and is monitored by integrated computers is the:

A) instrument cluster.
B) work cell.
C) conveyor belt.
D) workstation.
سؤال
In a chemiluminescent analyzer system:

A) the excitation event that produces light is caused by a chemical reaction.
B) reagents are impregnated on films or filter paper strips of the slide.
C) the intensity of emitted light is directly proportional to the concentration of analyte.
D) diffuse reflected light is measured.
سؤال
The 80% rule of thumb that is used by some laboratories in guiding decisions regarding automating a complete laboratory to achieve cost reduction means:

A) if 80% of the specimen handling situation can be standardized and automated,then the investment in automation is justified.
B) that 80% of a laboratory's income will have to be dedicated to the implementation and upkeep of all automated processes.
C) if 80% of the individuals involved in the laboratory process agree,then automation will be implemented.
D) approximately 80% of individuals will have to give up their jobs.
سؤال
The most significant improvement in the quality of laboratory test results can be attributed to the ability of automation to:

A) reduce costs.
B) increase work capacity.
C) reduce error and variability caused by manual manipulation.
D) reduce sample volumes and laboratory involvement.
سؤال
The steps involved in the completion of an automated analytical process are collectively referred to as:

A) aliquots.
B) unit operations.
C) instrument clusters.
D) sequential analysis.
سؤال
In an automated laboratory workstation with a bidirectional laboratory information system interface,the laboratory's bar-coded label on an individual specimen contains information regarding identification and the laboratory tests requested.The processor reads the bar code and the analyzer performs the analyses.Where can a preanalytical error still occur in this automated system?

A) When the instrument's calibration is not successful
B) When the quality control materials are out of the recommended range
C) When a data entry person enters incorrect laboratory test requests
D) When the equipment operator ignores equipment diagnostic messages
فتح الحزمة
قم بالتسجيل لفتح البطاقات في هذه المجموعة!
Unlock Deck
Unlock Deck
1/15
auto play flashcards
العب
simple tutorial
ملء الشاشة (f)
exit full mode
Deck 16: Automation
1
The type of automated analyzer in which the operator is able to use in-house reagents or reagents purchased from suppliers different from the analyzer's manufacturer is referred to as a(n):

A) discrete analyzer.
B) random-access analyzer.
C) open analyzer system.
D) closed analyzer system.
open analyzer system.
2
Your laboratory has purchased a discrete processing analyzer that is dedicated to performing only thyroid hormone analysis.Which type of pipetting system will likely be in place in this instrument?

A) Positive liquid displacement fixed-volume
B) Air displacement selectable-volume
C) Positive liquid displacement selectable-volume
D) Air displacement fixed-volume
Positive liquid displacement fixed-volume
3
The most commonly used method of reducing carryover in most automated analytical systems is to use:

A) extensive system flushing between specimens and sample probe washing.
B) plastic reagent containers with closed lids.
C) disposable cuvets.
D) a specific measurement approach that eliminates any possibility of carryover.
extensive system flushing between specimens and sample probe washing.
4
If programmed appropriately,a multiple-channel analyzer could assess samples as a single-channel analyzer.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
5
What component causes carryover of the analyte in a discrete analytical system that uses disposable reaction vessels and measuring cuvets?

A) The point at which sequential samples pass through
B) The pipetting system
C) The peristaltic pump
D) The reagent delivery system
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
6
A tray of clear sample cups sits on a laboratory bench top for an extended period of time.Which one of the following analytes might be affected?

A) Protein
B) Cholesterol
C) Glucose
D) Bilirubin
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
7
Using whole blood as the specimen of choice in an automated analytical system:

A) essentially eliminates specimen preparation time.
B) allows the operator to use a secondary tube for analysis.
C) keeps the specimen from undergoing degradation.
D) allows for the avoidance of carryover.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
8
An ion-selective electrode system for electrolyte analysis that uses a peristaltic pump to move liquid reagents and samples through a system of tubing is called a _____ analyzer.

A) random access
B) batch
C) discrete
D) continuous flow
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
9
Specimen bar coding has allowed for the ability of a system operator to place specimens in random sequence,thereby eliminating the need for specimen load lists.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
10
The point of interaction between the instrument operator and an automated analyzer that is dedicated to a defined task and is monitored by integrated computers is the:

A) instrument cluster.
B) work cell.
C) conveyor belt.
D) workstation.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
11
In a chemiluminescent analyzer system:

A) the excitation event that produces light is caused by a chemical reaction.
B) reagents are impregnated on films or filter paper strips of the slide.
C) the intensity of emitted light is directly proportional to the concentration of analyte.
D) diffuse reflected light is measured.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
12
The 80% rule of thumb that is used by some laboratories in guiding decisions regarding automating a complete laboratory to achieve cost reduction means:

A) if 80% of the specimen handling situation can be standardized and automated,then the investment in automation is justified.
B) that 80% of a laboratory's income will have to be dedicated to the implementation and upkeep of all automated processes.
C) if 80% of the individuals involved in the laboratory process agree,then automation will be implemented.
D) approximately 80% of individuals will have to give up their jobs.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
13
The most significant improvement in the quality of laboratory test results can be attributed to the ability of automation to:

A) reduce costs.
B) increase work capacity.
C) reduce error and variability caused by manual manipulation.
D) reduce sample volumes and laboratory involvement.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
14
The steps involved in the completion of an automated analytical process are collectively referred to as:

A) aliquots.
B) unit operations.
C) instrument clusters.
D) sequential analysis.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
15
In an automated laboratory workstation with a bidirectional laboratory information system interface,the laboratory's bar-coded label on an individual specimen contains information regarding identification and the laboratory tests requested.The processor reads the bar code and the analyzer performs the analyses.Where can a preanalytical error still occur in this automated system?

A) When the instrument's calibration is not successful
B) When the quality control materials are out of the recommended range
C) When a data entry person enters incorrect laboratory test requests
D) When the equipment operator ignores equipment diagnostic messages
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.
فتح الحزمة
k this deck
locked card icon
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 15 في هذه المجموعة.