Deck 40: Quality Assessment in the Hematology Laboratory
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Question
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/41
Play
Full screen (f)
Deck 40: Quality Assessment in the Hematology Laboratory
1
Regarding the evaluation of proficiency testing results, which of the following sets the tolerance limit for a given hematology procedure (e.g., hemoglobin)?
A) The Joint Commission
B) CLIA '88
C) College of American Pathologists
D) COLA
A) The Joint Commission
B) CLIA '88
C) College of American Pathologists
D) COLA
CLIA '88
2
Which of the following statistical measurements can be used to analyze a new instrument's precision?
A)F-test
B)Linear regression
C)Calculated SD
D)T-test
A)F-test
B)Linear regression
C)Calculated SD
D)T-test
F-test
3
On average, how often should proficiency testing be performed in the laboratory?
A)Once a month
B)Three times a year
C)Once a year
D)Every day
A)Once a month
B)Three times a year
C)Once a year
D)Every day
Three times a year
4
The following numbers are on a data set for a new lot of QC material: 23, 25, 19, 27, 20, 22, 24, and 26.What are the mean and standard deviation?
A)27.45; 2.82
B)23.25; 3.15
C)22.47; 1.45
D)23.25; 2.82
A)27.45; 2.82
B)23.25; 3.15
C)22.47; 1.45
D)23.25; 2.82
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
5
Quality control is run on the Sysmex XE-2100 for the morning shift.The low control is outside the 2SD limit for the WBC parameter.The technologist investigates the quality control material for low volume, checks reagents, and checks the previous QC charts.There are no discrepancies.What are the technologist's next steps in investigating this finding?
A)Report out the QC results and run patient samples.
B)Notify the supervisor and calibrate the instrument.
C)Notify the company and perform periodic maintenance.
D)Ignore the QC results and run patient samples.
A)Report out the QC results and run patient samples.
B)Notify the supervisor and calibrate the instrument.
C)Notify the company and perform periodic maintenance.
D)Ignore the QC results and run patient samples.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
6
In interpreting a patient's CBC results, the clinical laboratory professional notes that the platelet count is 25 x 109/L.Which of the following could be associated with this finding?
A)Observation of RBC fragments on the blood smear
B)Presence of cryoglobulins
C)Observation of platelet clumps at the feather edge of the blood smear
D)Presence of intracellular parasites such as malaria
A)Observation of RBC fragments on the blood smear
B)Presence of cryoglobulins
C)Observation of platelet clumps at the feather edge of the blood smear
D)Presence of intracellular parasites such as malaria
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
7
A sample collected for a prothrombin time is delivered to the laboratory 4 hours after collection.Which source of error does this represent?
A) Pre-examination
B) Examination
C) Post-examination Correct
A) Pre-examination
B) Examination
C) Post-examination Correct
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
8
Which of the following represents an appropriate step to meet universal precautions?
A) Check the sample source before using protective equipment
B) Require HIV immunization for all healthcare workers
C) Use mechanical pipeting devices to deliver all sample aliquots
D) Dispose of capped needles in biohazard trash bags
A) Check the sample source before using protective equipment
B) Require HIV immunization for all healthcare workers
C) Use mechanical pipeting devices to deliver all sample aliquots
D) Dispose of capped needles in biohazard trash bags
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
9
Implementation of a quality assessment program in laboratories is mandated by:
A)OSHA.
B)JCAHO.
C)CLIA 88.
D)CAP.
A)OSHA.
B)JCAHO.
C)CLIA 88.
D)CAP.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
10
Improper patient identification is considered a violation of which of the following?
A)Pre-examination QA
B)Examination QA
C)Post-examination QA
D)All of the above
A)Pre-examination QA
B)Examination QA
C)Post-examination QA
D)All of the above
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
11
The treatment of everything in the laboratory as infectious is defined as:
A)Proficiency testing.
B)Quality control.
C)Reference interval.
D)Universal precautions.
A)Proficiency testing.
B)Quality control.
C)Reference interval.
D)Universal precautions.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
12
Which of the following is an acceptable means for competency testing?
A)Direct observation checklists
B)Random assignment of proficiency samples
C)Blinded pre-analyzed samples
D)All of the above
A)Direct observation checklists
B)Random assignment of proficiency samples
C)Blinded pre-analyzed samples
D)All of the above
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
13
The regulatory body responsible for establishing proper guidelines for biohazardous waste material removal is:
A)OSHA.
B)CAP.
C)JCAHO.
D)MSDS.
A)OSHA.
B)CAP.
C)JCAHO.
D)MSDS.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
14
Who is responsible for training clinical laboratory professionals regarding the use of hazardous chemicals?
A) Human resources department
B) Infection control department
C) Laboratory manager
D) Biotechnology specialist
A) Human resources department
B) Infection control department
C) Laboratory manager
D) Biotechnology specialist
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
15
Why is it recommended that only 5-10 subjects be tested per day during a reference interval study?
A) Ensure that all subjects have properly prepared for the blood draw
B) Minimize effect of increased workload on laboratory operations
C) Increase ability to recruit sufficient numbers of subjects
D) Reduce introduction of an instrument or reagent bias
A) Ensure that all subjects have properly prepared for the blood draw
B) Minimize effect of increased workload on laboratory operations
C) Increase ability to recruit sufficient numbers of subjects
D) Reduce introduction of an instrument or reagent bias
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
16
A patient's automated reticulocyte count was 203 x 10⁹/L.Which of the following could account for this result?
A) Heinz bodies
B) Platelet satellitism
C) Leukocytosis
D) Fibrin strands
A) Heinz bodies
B) Platelet satellitism
C) Leukocytosis
D) Fibrin strands
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
17
Which of the following would cause the technologist to question the specimen integrity of a coagulation blood sample?
A)Hemolysis
B)Blood:anticoagulant ratio < 9:1
C)Blood:anticoagulant ratio > 9:1
D)All of the above
A)Hemolysis
B)Blood:anticoagulant ratio < 9:1
C)Blood:anticoagulant ratio > 9:1
D)All of the above
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
18
The following 20 hemoglobin results were collected for Level I control to determine its initial control limit (Results are in g/dL):
What are the 2s limits?
A) 5.8-6.4 g/dL
B) 5.8-6.5 g/dL
C) 5.5-6.7 g/dL
D) 5.4-6.7 g/dL
What are the 2s limits?
A) 5.8-6.4 g/dL
B) 5.8-6.5 g/dL
C) 5.5-6.7 g/dL
D) 5.4-6.7 g/dL
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
19
Which hematologic parameters would be affected by excessive lipemia in the blood sample?
A)MCV, MCH, MCHC
B)Hgb, MCH, MCHC
C)Hct, MCV, MCHC
D)WBC and differential
A)MCV, MCH, MCHC
B)Hgb, MCH, MCHC
C)Hct, MCV, MCHC
D)WBC and differential
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
20
The purpose of a control material is to:
A) Ensure the reliability of a given test procedure.
B) Determine if a patient result is normal, increased or decreased.
C) Establish the calibration factor for calculation of results.
D) Verify the reportable range for a given test procedure.
A) Ensure the reliability of a given test procedure.
B) Determine if a patient result is normal, increased or decreased.
C) Establish the calibration factor for calculation of results.
D) Verify the reportable range for a given test procedure.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
21
A clinical laboratory professional performed a procedure to verify the hemoglobin reportable range on the Cell-Dyn 4000.The highest level of linearity check material was 22.0 g/dL.This material was run several times with 21.7 g/dL, representing the mean.How should this finding be interpreted?
A)The result is within the tolerance limits, and the reportable range is verified.
B)The result exceeds the tolerance limits, and the procedure should be repeated with new check material.
C)The result exceeds the tolerance limits, and additional check material should be run in the upper limits of the reportable range.
D)The result exceeds the tolerance limits, and the verification of reportable range should be repeated after the instrument is recalibrated.
A)The result is within the tolerance limits, and the reportable range is verified.
B)The result exceeds the tolerance limits, and the procedure should be repeated with new check material.
C)The result exceeds the tolerance limits, and additional check material should be run in the upper limits of the reportable range.
D)The result exceeds the tolerance limits, and the verification of reportable range should be repeated after the instrument is recalibrated.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
22
When using moving averages to monitor an automated hematology instrument, the clinical laboratory professional notes that a moving average falls out of range, but the recently performed quality control was acceptable, what would this indicate?
A) The instrument should be recalibrated.
B) The range is incorrect for the moving average.
C) The patient samples for last 20 results were from a single location.
D) The clinical laboratory professional should open and run new control materials.
A) The instrument should be recalibrated.
B) The range is incorrect for the moving average.
C) The patient samples for last 20 results were from a single location.
D) The clinical laboratory professional should open and run new control materials.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
23
An instrument printout reveals the following: PT = 9.0 sec and APTT = 20.0 sec.What should the technologist do next?
A)Report patient results.
B)Check the sample for the presence of small clots.
C)Observe the sample for the presence of hemolysis.
D)Examine the collection tube for underfilling.
A)Report patient results.
B)Check the sample for the presence of small clots.
C)Observe the sample for the presence of hemolysis.
D)Examine the collection tube for underfilling.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
24
What is the first step in the selection of a new instrument?
A)Conduct a needs assessment survey.
B)Determine the cost per test.
C)Perform an in-house evaluation of the instrument.
D)Evaluate the quality control program offered.
A)Conduct a needs assessment survey.
B)Determine the cost per test.
C)Perform an in-house evaluation of the instrument.
D)Evaluate the quality control program offered.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
25
What are the components of a method evaluation? Explain the purpose of each.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
26
Moving averages can be used to monitor the instrument's performance in determining parameters of which cell line?
A)Leukocyte
B)Erythrocyte
C)Platelets
D)Reticulocytes
A)Leukocyte
B)Erythrocyte
C)Platelets
D)Reticulocytes
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
27
The following CBC results were obtained on an adult female:
Which is the most appropriate action?
A) Report the CBC results to the physician.
B) Repeat the CBC to confirm hemoglobin.
C) Examine peripheral blood smear for giant platelets.
D) Perform saline replacement for plasma and rerun.
Which is the most appropriate action?
A) Report the CBC results to the physician.
B) Repeat the CBC to confirm hemoglobin.
C) Examine peripheral blood smear for giant platelets.
D) Perform saline replacement for plasma and rerun.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
28
In reviewing a patient's APTT result, the clinical laboratory professional notes the presence of a delta flag.The patient's current APTT is 54 seconds, while the previous APTT (24 hrs ago) was 31 seconds (Note: QC was acceptable for both runs).Which is the most appropriate action?
A) Report the APTT since QC was acceptable.
B) Check the sample for the presence of clots.
C) Repeat APTT using a different methodology.
D) Perform a mixing study procedure on the sample.
A) Report the APTT since QC was acceptable.
B) Check the sample for the presence of clots.
C) Repeat APTT using a different methodology.
D) Perform a mixing study procedure on the sample.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
29
What is the goal of the Westgard multirule protocol for evaluating control results?
A) Reject all runs in which 1₂s control rule is violated.
B) Eliminate random error detection.
C) Minimize false rejection of a control run.
D) Increase analytical sensitivity of a control run.
A) Reject all runs in which 1₂s control rule is violated.
B) Eliminate random error detection.
C) Minimize false rejection of a control run.
D) Increase analytical sensitivity of a control run.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
30
What is the importance of documentation in a quality assessment program?
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
31
Which of the following represents the BEST method to assess a clinical laboratory professional's competency in performing leukocyte differentials?
A) Give the individual a 25-question written examination
B) Observe the individual during performance of a leukocyte differential
C) Assign the individual to complete the next CAP survey for hematology
D) Monitor the number of differentials that require pathology review for this individual
A) Give the individual a 25-question written examination
B) Observe the individual during performance of a leukocyte differential
C) Assign the individual to complete the next CAP survey for hematology
D) Monitor the number of differentials that require pathology review for this individual
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
32
An instrument printout reveals a hemoglobin of 15.0 g/dL, an RBC count of 3.0 x 1012/L, and a hematocrit of 30%.The MCHC is 50g/dL.A manual hematocrit is performed, and the technologist observes gross hemolysis and lipemia in the manual hematocrit tube.What is the technologist's next step in correcting the problem?
A)Re-spin the manual hematocrit.
B)Warm the sample and re-run it.
C)Call for a redraw.
D)Perform a plasma replacement.
A)Re-spin the manual hematocrit.
B)Warm the sample and re-run it.
C)Call for a redraw.
D)Perform a plasma replacement.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
33
Name one way that the laboratory verifies/validates the following:
A)Accuracy
B)Precision
C)Analytical sensitivity
D)Analytical specificity
A)Accuracy
B)Precision
C)Analytical sensitivity
D)Analytical specificity
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
34
Identify the critical information that must be on a material safety data sheet.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
35
Which of the following Westgard rule violations allows for patient samples to be run?
A)1-3s
B)1-2s
C)R-41s
D)4-1s
A)1-3s
B)1-2s
C)R-41s
D)4-1s
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
36
Given the following fibrinogen control limits: Level I
Level II
If the multirule control procedure for the fibrinogen assay is 13s, 22s, and R4s, how would you interpret the fibrinogen control results for the current run?
A) Acceptable
B) Unacceptable, 1?s violation
C) Unacceptable, 2?s violation
D) Unacceptable, R?s violation
Level II
If the multirule control procedure for the fibrinogen assay is 13s, 22s, and R4s, how would you interpret the fibrinogen control results for the current run?
A) Acceptable
B) Unacceptable, 1?s violation
C) Unacceptable, 2?s violation
D) Unacceptable, R?s violation
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
37
Which performance study in the method evaluation process determines the reportable range for the test method?
A) Precision study
B) Linearity study
C) Calibration study
D) Method comparison study
A) Precision study
B) Linearity study
C) Calibration study
D) Method comparison study
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
38
Patient samples are run on hematology analyzers to assess the instrument's:
A)Accuracy.
B)Precision.
C)Specificity.
D)Sensitivity.
A)Accuracy.
B)Precision.
C)Specificity.
D)Sensitivity.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
39
Which source of error would be detected by a delta check?
A)Depleted reagent supply
B)Improper calibration of the instrument
C)Deteriorating light source
D)Failure to
A)Depleted reagent supply
B)Improper calibration of the instrument
C)Deteriorating light source
D)Failure to
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
40
Given the following precision study results for an automated prothrombin time procedure:
Which is the appropriate interpretation?
A) Precision study results are acceptable.No random error was detected.
B) Precision study results are unacceptable.Instrument should be recalibrate and precision study repeated.
C) Precision study results are unacceptable.The sampling pipet should be replaced and precision study repeated.
D) Precision study results are unacceptable.Review results for outliers if present eliminate from data and recalculate CV.
Which is the appropriate interpretation?
A) Precision study results are acceptable.No random error was detected.
B) Precision study results are unacceptable.Instrument should be recalibrate and precision study repeated.
C) Precision study results are unacceptable.The sampling pipet should be replaced and precision study repeated.
D) Precision study results are unacceptable.Review results for outliers if present eliminate from data and recalculate CV.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck
41
Name the hematologic/hemostatic parameters that are affected by hemolysis, and offer a means by which to correct the problem.
Unlock Deck
Unlock for access to all 41 flashcards in this deck.
Unlock Deck
k this deck

