Deck 6: Quality Assurance and Calibration Methods

Full screen (f)
exit full mode
Question
The percent by mass octane is quantified for an organic sample via GC using decane as the internal standard.A GC vial containing 0.5023 grams organic sample and 0.2384 grams octane is prepared and analyzed,returning peak areas of 119,993 for octane and 83755 for decane.If a response factor of 2.50 from previous work analyzing octane with a decane internal standard is used,what is the percent by mass octane in the organic sample?
Use Space or
up arrow
down arrow
to flip the card.
Question
There are several types of precision involved with method validation.

A)instrument precision
B)final precision
C)intermediate precision
D)interlaboratory precision
E)intra-assay precision
Question
A____________________ is a sample that has been exposed to the site of sampling,containing all components except analyte,and it is taken through all steps of the analytical procedure.

A)method blank
B)reagent blank
C)field blank
D)spike blank
E)matrix blank
Question
A(n)________________________________________is a known amount of compound,different from the analyte,that is added to the unknown.

A)standard addition
B)internal standard
C)external standard
D)self-contained standard
E)known standard
Question
A set of five solutions are prepared by delivering 10 mL of unknown sample and increasing volumes of standard (0,5,10,15,and 20 mL)into 50-mL volumetric flasks.Analytical reagents are then added and diluted to volume.The x-intercept is −3.90 mM.What is the unknown concentration?

A)39.0 mM
B)156 mM
C)19.5 mM
D)0)390 mM
E)195 mM
Question
Which are TRUE statements regarding internal standards?
I The response factor is calculated by using a mixture containing known amounts of analyte and internal standard and by using detector response to the analyte and the internal standard.
II Internal standards are useful where instrument response or quantity of sample analyzed varies over time.
III Detector response is inversely proportional to concentration.
IV Detector response to analyte and internal standard is assumed to be constant over a range of concentrations.

A)I,II,and III
B)II,III,and IV
C)I and II
D)I and III
E)I,II,and IV
Question
The____________________ state(s)what steps will be taken and how they will be carried out.

A)assessment
B)calibration check
C)control charts
D)specifications
E)standard operating procedure
Question
Which of the following is(are)not a way to demonstrate the accuracy of a method?
I Determine the limit of detection for the method.
II Analysis of a certified reference material in a matrix similar to the sample.
III Compare the results between two or more different analytical methods.
IV Analyze a blank sample with a known amount of analyte in the same matrix as the sample.

A)Only IV
B)II and III
C)I and III
D)Only III
E)Only I
Question
____________________ is the concentration range over which there is measurable instrument response.

A)Accuracy range
B)Dynamic range
C)Linear range
D)Detection range
E)Range
Question
A____________________ is an analytical result that indicates the sample concentration is below a stated limit when the concentration is actually above the limit.

A)positive result
B)false negative
C)negative result
D)false positive
E)negative
Question
A solution is 2.0% by mass compound A and 3.5% by mass compound I.Separation via GC returns peak areas of 38234 for compound A and 23880 for compound I.Calculate the response factor.

A)0)664
B)2)03
C)1)50
D)0)491
E)1)17
Question
The detection limit for caffeine analysis must be calculated.Eight absorbance readings from a caffeine sample close to the estimated detection limit are 0.1239,0.1098,0.1182,0.1150,0.1219,0.1251,0.1199,and 0.1099.The average absorbance for eight blanks is 0.0182.The slope of a calibration curve is 0.59 /mM.What is the detection limit?
Question
Which statement regarding standard addition to a single solution is FALSE? A Analyte signal is measured after each standard addition.

A)The volume standard added must be kept small to minimize altering the matrix.
B)Detector response is corrected to take in account dilution before being plotted on the y-axis.
C)The absolute value of the y-intercept for the best fit line is the unknown analyte concentration.
D)Volume diluted standard is plotted on the x-axis.
Question
There are many studies required to validate an analytical method.Which of the studies below is not required for method validation?

A)method specificity
B)linearity
C)range
D)calibration
E)limit of detection
Question
A water sample collected downstream from a gold-mining operation returns a cyanide concentration of 6.18 μ\mu g/L CN.A 10-mL aliquot is spiked with 100 μ\mu L of 500.0 μ\mu g/L sodium cyanide and analyzed.Analysis returns a cyanide concentration of 10.5 μ\mu g/L CN.What is the percent recovery for the spike?
Question
Which statement regarding the multiple solutions with constant volume standard addition method is FALSE?

A)Equal volumes of unknown are pipetted into a set of volumetric flasks with identical volume.
B)Equal volumes of standard are pipetted into each volumetric flask.
C)Analytical reagents are added to each volumetric flask and diluted to volume.
D)Detector response versus diluted standard is plotted.
E)The absolute value of the x-intercept for the best-fit line is the diluted unknown concentration.
Question
A chemist is tasked with determining the concentration of nickel for a water sample drawn from a well next to a nickel mine.To analyze the sample using atomic absorbance spectroscopy (AAS),she prepares six samples (0-5)by adding 10.00 mL well water to a 50.00-mL volumetric flask,then adding an increasing volumes of 25.00-ppm nickel solution,diluting the resulting solution to 50.00 mL with 0.1 M HNO3.Following AAS analysis,a graph of absorbance versus ppm nickel standard gave a best-fit line y = 0.0428x + 0.0487.Calculate the ppm nickel in the well water sample.
Question
A____________________ is a sample containing all components except analyte,and it is taken through all steps of the analytical procedure.

A)method blank
B)reagent blank
C)field blank
D)spike blank
E)matrix blank
Question
The response factor,F,for the analysis of pentane with the internal standard decane is 3.5.Which statements below are INCORRECT for the analysis?
I The pentane concentration is 3.5 times greater than the decane concentration.
II The decane concentration is 3.5 times greater than the pentane concentration.
III Detector response to pentane is 3.5 times greater than detector response to decane.
IV Detector response to decane is 3.5 times greater than detector response to pentane.

A)I and III
B)II and III
C)II and IV
D)I and IV
E)None of these is incorrect.
Question
The____________________ is everything in the sample,except analyte.

A)blind sample
B)performance test sample
C)spike
D)matrix
E)assessment
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/20
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 6: Quality Assurance and Calibration Methods
1
The percent by mass octane is quantified for an organic sample via GC using decane as the internal standard.A GC vial containing 0.5023 grams organic sample and 0.2384 grams octane is prepared and analyzed,returning peak areas of 119,993 for octane and 83755 for decane.If a response factor of 2.50 from previous work analyzing octane with a decane internal standard is used,what is the percent by mass octane in the organic sample?
27.3% octane
2
There are several types of precision involved with method validation.

A)instrument precision
B)final precision
C)intermediate precision
D)interlaboratory precision
E)intra-assay precision
final precision
3
A____________________ is a sample that has been exposed to the site of sampling,containing all components except analyte,and it is taken through all steps of the analytical procedure.

A)method blank
B)reagent blank
C)field blank
D)spike blank
E)matrix blank
field blank
4
A(n)________________________________________is a known amount of compound,different from the analyte,that is added to the unknown.

A)standard addition
B)internal standard
C)external standard
D)self-contained standard
E)known standard
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
5
A set of five solutions are prepared by delivering 10 mL of unknown sample and increasing volumes of standard (0,5,10,15,and 20 mL)into 50-mL volumetric flasks.Analytical reagents are then added and diluted to volume.The x-intercept is −3.90 mM.What is the unknown concentration?

A)39.0 mM
B)156 mM
C)19.5 mM
D)0)390 mM
E)195 mM
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
6
Which are TRUE statements regarding internal standards?
I The response factor is calculated by using a mixture containing known amounts of analyte and internal standard and by using detector response to the analyte and the internal standard.
II Internal standards are useful where instrument response or quantity of sample analyzed varies over time.
III Detector response is inversely proportional to concentration.
IV Detector response to analyte and internal standard is assumed to be constant over a range of concentrations.

A)I,II,and III
B)II,III,and IV
C)I and II
D)I and III
E)I,II,and IV
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
7
The____________________ state(s)what steps will be taken and how they will be carried out.

A)assessment
B)calibration check
C)control charts
D)specifications
E)standard operating procedure
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
8
Which of the following is(are)not a way to demonstrate the accuracy of a method?
I Determine the limit of detection for the method.
II Analysis of a certified reference material in a matrix similar to the sample.
III Compare the results between two or more different analytical methods.
IV Analyze a blank sample with a known amount of analyte in the same matrix as the sample.

A)Only IV
B)II and III
C)I and III
D)Only III
E)Only I
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
9
____________________ is the concentration range over which there is measurable instrument response.

A)Accuracy range
B)Dynamic range
C)Linear range
D)Detection range
E)Range
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
10
A____________________ is an analytical result that indicates the sample concentration is below a stated limit when the concentration is actually above the limit.

A)positive result
B)false negative
C)negative result
D)false positive
E)negative
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
11
A solution is 2.0% by mass compound A and 3.5% by mass compound I.Separation via GC returns peak areas of 38234 for compound A and 23880 for compound I.Calculate the response factor.

A)0)664
B)2)03
C)1)50
D)0)491
E)1)17
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
12
The detection limit for caffeine analysis must be calculated.Eight absorbance readings from a caffeine sample close to the estimated detection limit are 0.1239,0.1098,0.1182,0.1150,0.1219,0.1251,0.1199,and 0.1099.The average absorbance for eight blanks is 0.0182.The slope of a calibration curve is 0.59 /mM.What is the detection limit?
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
13
Which statement regarding standard addition to a single solution is FALSE? A Analyte signal is measured after each standard addition.

A)The volume standard added must be kept small to minimize altering the matrix.
B)Detector response is corrected to take in account dilution before being plotted on the y-axis.
C)The absolute value of the y-intercept for the best fit line is the unknown analyte concentration.
D)Volume diluted standard is plotted on the x-axis.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
14
There are many studies required to validate an analytical method.Which of the studies below is not required for method validation?

A)method specificity
B)linearity
C)range
D)calibration
E)limit of detection
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
15
A water sample collected downstream from a gold-mining operation returns a cyanide concentration of 6.18 μ\mu g/L CN.A 10-mL aliquot is spiked with 100 μ\mu L of 500.0 μ\mu g/L sodium cyanide and analyzed.Analysis returns a cyanide concentration of 10.5 μ\mu g/L CN.What is the percent recovery for the spike?
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
16
Which statement regarding the multiple solutions with constant volume standard addition method is FALSE?

A)Equal volumes of unknown are pipetted into a set of volumetric flasks with identical volume.
B)Equal volumes of standard are pipetted into each volumetric flask.
C)Analytical reagents are added to each volumetric flask and diluted to volume.
D)Detector response versus diluted standard is plotted.
E)The absolute value of the x-intercept for the best-fit line is the diluted unknown concentration.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
17
A chemist is tasked with determining the concentration of nickel for a water sample drawn from a well next to a nickel mine.To analyze the sample using atomic absorbance spectroscopy (AAS),she prepares six samples (0-5)by adding 10.00 mL well water to a 50.00-mL volumetric flask,then adding an increasing volumes of 25.00-ppm nickel solution,diluting the resulting solution to 50.00 mL with 0.1 M HNO3.Following AAS analysis,a graph of absorbance versus ppm nickel standard gave a best-fit line y = 0.0428x + 0.0487.Calculate the ppm nickel in the well water sample.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
18
A____________________ is a sample containing all components except analyte,and it is taken through all steps of the analytical procedure.

A)method blank
B)reagent blank
C)field blank
D)spike blank
E)matrix blank
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
19
The response factor,F,for the analysis of pentane with the internal standard decane is 3.5.Which statements below are INCORRECT for the analysis?
I The pentane concentration is 3.5 times greater than the decane concentration.
II The decane concentration is 3.5 times greater than the pentane concentration.
III Detector response to pentane is 3.5 times greater than detector response to decane.
IV Detector response to decane is 3.5 times greater than detector response to pentane.

A)I and III
B)II and III
C)II and IV
D)I and IV
E)None of these is incorrect.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
20
The____________________ is everything in the sample,except analyte.

A)blind sample
B)performance test sample
C)spike
D)matrix
E)assessment
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 20 flashcards in this deck.